Late and Spontaneous Extrusion of Xen45 Glaucoma Gel Stent

Silvia Babighian*, MD, PhD, Tommaso Candian1, MD, Alessandro Galan1, MD

 

1. Department of Ophthalmology, Ospedale Sant’ Antonio, Azienda Ospedaliera, Padova - Italy.


Corresponding Author: Silvia Babighian, Silvia Babighian, MD, PhD, Department of Ophthalmology, Ospedale Sant’Antonio, Azienda Ospedaliera, Via Facciolati 71, 35127 Padova, Italy.

Copy Right: © 2022 Silvia Babighian, This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


Received Date: October 17, 2022

Published Date: November 01, 2022


Late and Spontaneous Extrusion of Xen45 Glaucoma Gel Stent

Case Report

In October 2016, a 84-year-old woman with primary open-angle glaucoma and atrophic maculopathy, bilaterally pseudophakic, underwent Xen45® Glaucoma Gel Stent (XEN-GGS; Allergan, Irvine, CA, USA) implantation in her left eye for an increase of intraocular pressure (IOP) to 23 mmHg despite antiglaucoma medical therapy. Surgery was performed in our hospital, under local anaesthesia according to the manufacturer’s instructions. An injection of 0.1 ml mitomycin C (MMC) 0.2 mg/ml was performed in the subconjunctival space to reduce postoperative scarring at the filtration site. Clear corneal incisions were created and the anterior chamber (AC) was filled with viscoelastic. The device, preloaded in the XEN Injector, was inserted ab interno towards the superonasal quadrant, placed to anterior trabecular meshwork, advanced through the sclera and released under the conjunctiva. The corrected placement was confirmed by intraoperative gonioscopy. The viscoelastic was washed out creating a subconjunctival bleb and confirming patency of the device. At the end of surgery, the recommended placement of XEN-GGS 1-2-3 mm (AC, sclera, and subconjunctival area) was confirmed and surgery completed.

Postoperative treatment consisted of combination netilmicin and dexamethasone (Netildex®) eyedrops instilled four times a day for one month followed by a slow taper over the second month. All glaucoma medications were discontinued on the day of surgery. Postoperative visits took place on day 1, day 7, day 21, 2 months and 4 months to continue follow-up every 6 months. The subsequent follow-ups were uneventful, with good IOP control between 8 and 14 mmHg. The filtering bleb appeared diffuse, broad and low elevated.

Four years later, in November 2020, the patient presented to the emergency department complaining of foreign body sensation in her left eye, lasting for 2 days. On presentation, the visual acuity was stable compared to baseline (left eye: 20/32; right eye: 20/50). The slit lamp examination revealed an erosion of the conjunctiva with exposure of the distal tip of XEN-GGS, exhibiting clear leakage (Fig.1 A). IOP was 8 mmHg, AC well formed and no choroidal detachment. No signs of ocular inflammation or infection were observed. XEN-GGS was carefully extracted with forceps, without rupture, at the slit lamp (Fig. 1B). After that, we treated the patient with local antibiotic 4 times/day for 7 days. Two days later the patient reported that the foreign body sensation had disappeared and the IOP remained at a normal value (10 mmHg). The patient was followed weekly during the next month with a quiet eye and a IOP in the low teens without antiglaucoma medications (Fig. 2).

 

Several factors may have predisposed to the XEN-GGS extrusion in our patient.

Firstly, as previously reported (Lim et al. 2018), low elevated blebs are at higher risk of leakage, due to the friction of the implant against the overlying conjunctiva, which can cause conjunctival erosion also leading to serious complications as blebitis or endophthalmitis.

Other possible risk factors of conjunctival erosion are the late effect of MMC on the tissue and the prolonged use of glaucoma eyedrops leading to chronic local inflammation as well as thinning of the conjunctiva (Arnljots et al 2019). Furthermore, it is well recognized that conjunctival tissue degenerates in the elderly (Abdel-Khalek et al. 1978). All these factors may have contributed to XEN-GGS extrusion in our old patient. Moreover, stopping of all topical therapy as postoperative steroids and antiglaucoma medications may have caused drying of the thin conjunctiva inducing irritation; this in turn led to the patient rubbing the eye, resulting in conjunctival erosion and exposure of XEN-GGS. In our report no complications worthy of note were detected and IOP was normal without antiglaucoma medication in according with the hypothesis that even if the entire XEN-GGS dislocates there is a possibility that the channel between the AC and subconjunctiva is maintained with achievement of control of IOP.

XEN-GGS is effective for lowering IOP, serves to reduce antiglaucoma medications, and may save surgical time compared to conventional glaucoma surgery. The correct positioning of the implant is mandatory for preventing complications (Lenzhofer et al 2019). This case report demonstrates that positioning XEN-GGS in or under Tenon layer rather than subconjunctivally, guarantees more stability of the device, reduces incidence of extrusion and is highly recommended in the elderly. To our knowledge this is the first report of spontaneous extrusion of XEN-GGS after four years.

 

Patient Consent

Informed written consent for publication has been obtained from the patient. This report does not contain any personal information that could lead to the identification of the patient.

 

Declaration of conflicting interest

The author(s) declared no potential conflicts of interest for the research, authorship, and/or publication of this article.


Funding

The author(s) received no financial support for the research, authorship, and/or publication of this article.

 

References

1. Lim R, Lim KS (2018): Xen implant-related endophthalmitis. Ophthalmology 125(2):209.

2. Arnljots TS, Economou MA (2019): Xen Gel Stent exposure 7 mo after primary implantation: a case report. Int J Ophthalmol 12(4):689-691.

3. Abdel-Khalek LM, Williamson J, Lee WR (1978): Morphological changes in the human conjunctival epithelium. I. In the normal elderly population. Br J Ophthalmol 62:792–799.

4. Lenzhofer M, Strohmaier C, Sperl P, Hohensinn M, Hitzl W, Steiner V, Moussa S, Motloch K, Baca B, Krall E, Reitsamer HA (2019): Effect of the outer stent position on efficacy after minimally invasive transscleral glaucoma gel stent implantation. Acta Ophthalmol 97(8):e1105-e1111.

antarmuka fokus mahjong daya pengguna aktifaws grid serasi mahjong dasar tahapan terjagaaws jejak mekanisme mahjong arah fase lanjutanaws kajian wild berantai mahjong interaktif analitisaws kesesuaian persentase layanan mahjong seluler lanceraws pendalaman persentase mahjong gerak wild mutakhircorak langka mahjong tumbuh perlahan berubahgerak mahjong adaptasi mekanisme pemakai sekarangnalar scatter mahjong malam putaran ekstratempo mahjong kaitan mekanisme keadaan terkinialur permainan mahjong cepat scatter wilddalam hitungan detik scatter wild mahjongmenyatukan naluri pola scatter hitam mahjongmomen mahjong permainan berbalik arahmomen singkat mahjong dinamika permainanperpaduan insting pola scatter hitam momentperubahan drastis mahjong ways scatter wildscatter wild mahjong datang polasekejap berubah scatter wild mahjong wayssensasi baru mahjong lebih scatter wildenergi scatter emas irama reel mahjongevolusi reel mahjong balutan mistisintervensi cepat emas momentum lamakemunculan mendadak naga emas mahjongketika scatter naga emas aktif mahjongnaga emas muncul arah spin mahjongnaga emas ritme mahjong ways berubahrahasia rtp tinggi balik scatter hitamsaat scatter naga emas alih irama reelscatter hitam kunci lonjakan rtp mahjonge5 scatter wild memberikan sentuhan baru di setiap spin mahjong ways 2e5 scatter wild menghidupkan suasana permainan mahjong ways 2e5 scatter wild mengubah pola permainan mahjong ways 2 secara signifikane5 setiap putaran mahjong ways 2 terasa berbeda dengan scatter wilde5 strategi adaptif berbasis analisis rtp hariane5 strategi berbasis data dan algoritma untuk analisis momentume5 strategi berkembang berkat data rtp hariane5 strategi memahami algoritma untuk mengidentifikasi momentum ideale5 strategi membaca pola algoritma demi menangkap momentum optimale5 strategi modern mengandalkan evaluasi rtp hariane5 strategi responsif dengan dukungan evaluasi rtp hariane5 strategi terukur dengan analisis rtp hariane5 struktur scatter dan wild terlihat jelas berkat analisis sistem moderne5 tanpa disadari kombinasi ini sering mengarah ke scatter di mahjong wins 3e5 teknik evaluasi algoritma untuk mendapatkan momentum yang tepate5 teknik observasi sistem untuk analisis momentum yang lebih presisie5 terungkap formasi ini sering jadi awal munculnya scatter di mahjong wins 3e5 transformasi digital rtp live berkat artificial intelligence inovatife5 transformasi ritme mahjong ways 2 dipicu oleh kekuatan scatter wilde5 wajib tahu pola tersembunyi ini sering menghasilkan scattere5 applee5 bananae5 candye5 doge5 eaglee5 falcone5 geminie5 horsee5 indiae5 japananalisa pola mahjong ways rutinanalisis kinerja heuristik variansi gameanalisis pola mahjong ways hariananalisis pola mahjong ways kebiasaanera baru mahjong wins bonus optimalgebrakan bonus mahjong wins mekanisme efisieninsight pola mahjong ways rutinkajian pola mahjong ways rutinkomparasi heuristik variansi game digitalledakan bonus mahjong wins sistem efektifmahjong wins bonus sistem generasi baruobservasi pola mahjong ways harianpendekatan algoritma heuristik variansi gameperbandingan model heuristik variansi gamerahasia bonus mahjong wins sistem cerdasrangkuman pola mahjong ways harianringkasan pola mahjong ways harianstudi pola mahjong ways hariantinjauan heuristik variansi game digitaltinjauan pola mahjong ways harianalur sombol mahjong kemunculan scatterdari rtp mahjong bermain lebih efektifjejak scatter mahjong putaran tenangkejutan scatter wild simbol mahjong arahkemunculan simbol ganda membuat mahjongketika grid mahjong scatter semakin dekatketika rtp mahjong pola mulai lebih jelasketika scatter wild ritme simbol mahjongketika scatter wild titik sesi mahjong waysketika susunan simbol mahjong ritme scattermemahami rtp mahjong cara bermain lebihpergerakan simbol mahjong scatter wildpergeseran mahjong ketika scatter hadirsaat rtp mahjong terbaca baik strategisaat scatter hadir simbol mahjong bergeserscatter wild dinamika simbol mahjongstabilitas putaran mahjong pola scattersusunan baru reel mahjong scatter emassusunan mahjong wins mengandung scattersusunan simbol mahjong diam pola scatterrm menguak keunikan mahjong wins sudut pandang teknisrm cara memahami pergerakan mahjong ways tenaga ekstrarm mahjong wins standar baru industri hiburan digitalrm rahasia ketahanan mahjong ways eksis gempuran gamerm pentingnya memahami transisi level mahjong wins mendalamrm strategi mengatur tempo mahjong ways kendali permainanrm peran kecerdasan buatan mekanisme mahjong wins adilrm alasan keberhasilan mahjong ways mencuri perhatian analisrm mempelajari struktur dasar mahjong wins efisiensi putaranrm inovasi desain mahjong ways kesan bermain responsifrm teknik observasi mahjong wins jarang dibahas dampakrm cara mempertahankan fokus dinamika mahjong ways cepatrm eksplorasi fitur tersembunyi mahjong wins ritme terbaikrm mahjong ways integrasi teknologi modern keamanan nyamanrm analisis faktor pendukung mahjong wins digemari generasirm langkah efektif menyesuaikan perubahan sistem mahjong waysrm mengintip proses pengembangan mahjong wins kualitas penggunarm analisis data membantu membaca arah mahjong waysrm menemukan titik temu insting logika mahjong winsrm transformasi besar mahjong ways menghadirkan tantangan menarikmengungkap simbol langka nasib drastismisteri besar kombinasi simbol langkamisteri simbol langka keberuntungan besarsimbol langka misterius ubah hiduprahasia simbol langka nasib cepattransformasi bonus mahjong wins sistem efektifmahjong wins suguhkan bonus sistem modernsuguhan bonus efisien mahjong winsefektivitas sistem bonus mahjong winsmahjong wins hadirkan bonus sistem optimaloke76cincinbetaqua365slot gacorstc76samurai76TOBA1131samurai76 login