Clinical Accuracy of Autorefraction in Adult Ethiopian Population

Clinical Accuracy of Autorefraction in Adult Ethiopian Population

Robel H. Ababiya *1, Samuel D. Shibeshi 2, Tsion H. Ababiya 3, Amanuel T. Tekle 4, Tsegasillase W/Yohannes 5, Bethelhem M. Terrefe 6, Bemnet A.Tamene 7


1. MD, Saint Paul's Hospital Millennium Medical College

2. MD, MPH, Addis Ababa University

3. MD, Internist, Wolayita Sodo University Hospital

4. MD, Asrade Zewde Primary Hospital

5. DDM, MPH, Nahom Speciality Dental clinic

6. Opthometrist, WGGA Eye Centre.

7. MD, Family Guidance Association.


Corresponding Author: Robel H. Ababiya, MD, Saint Paul's Hospital Millennium Medical College.

Copy Right: © 2023 Robel H. Ababiya, This is an open access article distributed under the Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.


Received Date: May 23, 2023

Published Date: June 01, 2023

 

Abstract

Purpose: The purpose of this study was to compare the accuracy of the Zeiss Visuref 100 Autorefractor using subjective refraction as a reference

Methods: one hundred ten patients aged 16 to 74 years with total eye of 220 underwent autorefraction followed by subjective refraction on the same day by a single clinician. Both eyes of each participants were included in the analysis.

Results: A clinically significant spherical equivalent difference was observed between the autorefraction and subjective refraction. Bland-Altman analysis revealed that the 95% limits of agreement between the autorefraction and subjective refraction range between -1.30 to +1.50 diopter.

Conclusions: Although autorefractor is a useful clinical tool as a starting point in refraction due to its clinically significant difference with subjective refraction it doesn’t replace the later and therefore shouldn’t be used alone.

 


Clinical Accuracy of Autorefraction in Adult Ethiopian Population

Introduction

Refractive errors are the most common ocular problem affecting all age groups. They are regarded as issues with public health. According to a recent WHO research, refractive errors account for 43% of visual impairments and are the second cause of visual loss worldwide as 43% of visual impairments are attributed to refractive errors.[1] In sub-Saharan Africa, the prevalence of refractive error and visual impairment due to refractive error was 12.6% and 3.4%, respectively.

 

Subjective refraction, retinoscopy, and autorefraction can be employed to evaluate the refractive error. Although there are alternatives to autorefraction, they may be time-consuming and occasionally wrong. In large-scale vision screening programs, autorefraction could be a beneficial component, particularly in resource-constrained areas like Ethiopia. Autorefraction enables health professionals to easily assess refractive error.[2]

 

Autorefractors provide the speed and precision needed for quick patient assessments in high-volume tertiary eye care settings. It could be a helpful element of extensive vision screening programs in public health outreach efforts, particularly in countries with low resources. [2]

 

About the accuracy of autorefractors in adults, there is a paucity of information worldwide. As far as the researcher is aware, there are no statistics on the accuracy of autorefractors in our nation. This research was performed to evaluate the reliability of autorefractors relative to subjective refraction. In this study, the Zeiss Visuref 100 Autorefractor's ability to provide a reasonably accurate estimate of refractive error was assessed.


Materials and Methods

Ethics

Ethical approval was obtained from institutional review boards at the St. Paul’s hospital millennium medical college. Written informed consent was obtained from all participants.

 

Study design

This was a cross-sectional study whereby 110 patients attending ophthalmology department at St. Paul’s hospital millennium medical college, were enrolled. Subjects were included only if they had vision 6/60 or better. All of the refraction procedures were performed under noncycloplegic conditions. Autorefraction was performed first, followed by subjective refraction. the Zeiss Visuref 100 Autorefractor was used since it was readily aviliable.  If the patient’s previous prescription was known it was used as the starting point for subjective refraction. Subjective refraction was conducted with trial frame with vertex distance kept to a minimum. Cylindrical power and axis were refined using a 0.25-diopter Jackson cross cylinder. Refraction of each eye was performed monocularly followed by binocular balancing. The highest plus sphere and the lowest minus results that provided the best acuity and comfortable vision were considered as the end point. Subjective refraction was recorded to the closest 0.25 diopter. To minimize the risk of bias, autorefraction was conducted by the same examiner. The autorefractor was calibrated daily according to the manufacturer’s guidelines. Participants were instructed to look at the internal fixation target.

                                                                         

Data analysis

Refraction measurements were transformed using Fourier vector decomposition to spherical equivalent and Jackson cross cylinder scalar values using the following formulas, where S = sphere, C = cylinder.

Bland-Altman plots of the spherical equivalent was used to demonstrate the limits of agreement between autorefraction and subjective refraction values, whereas mean differences were assessed using SPSS v26. The diagnostic accuracy of the autorefractor was assessed assuming subjective refraction as the reference standard. Subjective refraction was chosen as the reference standard because it is currently the test generally used to dispense eyeglassesses.                       


Results

In October 2022, study participants signed up. Both tests were completed on a total of 220 eyes from 110 individuals. No adverse events occurred during testing. Eyes with the best corrected visual acuity of 6/60 or worse were excluded. Approximately 50.1 % of participants were female and the average age of participants was 48 years(range 16 to 74 years). Subjectively assessed mean spherical equivalent refractive errors ranged from -20.00 to +4.00.                    


Analysis

Bland-Alman plots comparing measurements of spherical equivalent between autorefractor and subjective refraction are shown in Fig. Significant differences were observed in spherical equivalent findings of the autorefraction as compared to the subjective refraction. The 95% limits of agreement between the autorefraction and subjective refraction were -1.30 to +1.50 diopter(Fig). The autorefractor produced spherical equivalent estimates that were on average more myopic than subjective refraction(Table).

 

Discussion

In this investigation, the autorefractor produced slightly more myopic measurements of spherical equivalent compared to subjective refraction. The limits of agreement measured in this study indicate that 95% of the time, a measurement from an autorefractor will be within approximately 2.80 diopters of the measurement obtained from subjective refraction. Prior published studies have reported limits of agreement within 0.78-1.00 diopter.[3-5] However, these previous studies were done in a relatively low degree of refractive error therefore this might be one reason why the limit of agreement is low.

 

It's unclear what role autorefraction plays in adult vision screening programs. Unlike children where refractive error represents the most common cause of visual impairment in adults, other diseases like cataracts and diabetic retinopathy play a major role. Therefore, autorefraction may not be helpful in those diseases. Complete eye examinations with a slit lamp would be ideal to be able to capture these pathologies. Less expensive instruments like the pinhole occluder are sensitive and specific for the detection of refractive error, so it may not be necessary to include an expensive electronic device in screening programs [6,7]. On the other hand, older adults sometimes have difficulty navigating the pinhole test, whereas the autorefractor does not require much cooperation to provide a result.

 

The results of the present study suggests that autorefractors don't reach a level of diagnostic accuracy since there is a clinically significant difference when compared to subjective refraction but it can still be used as a starting point for patients who have both autorefraction and subjective refraction.

Even though we used subjective refraction as the benchmark in this investigation, it is not a perfect comparator[8]. For example, a study that compared subjective refraction performed by two different clinicians found a 95% limit of agreement of 0.78 D [3-5]. Moreover, subjective refraction required the cooperation and participation of the test taker.

 

Strengths of the study include a relatively large sample of eyes, and an older population not frequently studied in similar studies. Moreover, unlike previous studies which used a narrow range of refractive errors in their study, we included all patients who fulfilled the inclusion criteria.

It must be acknowledged that this study has some limitations. First, we concur with previous authors that using subjective refraction as a reference has potential drawbacks because measurements may vary depending on whether one or multiple medical professionals perform the subjective refraction. To reduce healthcare provider bias, we assessed subjective refraction by a single practitioner. The autorefractors were old versions, and we lacked access to modern models, which is another drawback.

 

Third, the subjective refraction may have been biased because the researcher was aware of the participant's prior prescription. However, the researcher made sure to follow a uniform procedure for every participant, with the goal being to get the greatest possible acuity and comfortable vision with the new prescription as the endpoint.

In conclusion, the 95% limit of agreement between the subjective refraction and objective refraction doesn't seem sufficient for the sole use of autorefraction in refractive error screening in community outreach programs. However, autorefraction may still be used as a quick method to identify patients who would benefit from spectacles, or as a starting point for a patient to compare several pairs of spectacles.


References

  1. Pascolini D., Mariotti S.P. Global estimates of visual impairment: 2010. Br J Ophthalmol. 2012;96(5):614–618. [PubMed] [Google Scholar] [Ref list]
  2. Pons J. Improving patient flow through an eye clinic. Community Eye Health. 2012;25(78):31-3.
  3. Kinge, B., Midelfart, A., & Jacobsen, G. (1996). Clinical evaluation of the Allergan Humphrey 500 autorefractor and the Nidek AR-1000 autorefractor. British Journal of Ophthalmology, 80(1), 35–39. https://doi.org/10.1136/bjo.80.1.35
  4. Pesudovs, K., & Weisinger, H. S. (2004). A Comparison of Autorefractor Performance. Optometry and Vision Science, 81(7), 554–558. https://doi.org/10.1097/00006324-200407000-00018
  5. Pesudovs, K., Parker, K. E., Cheng, H., & Applegate, R. A. (2007). The Precision of Wavefront Refraction Compared to Subjective Refraction and Autorefraction. Optometry and Vision Science, 84(5), 387–392. https://doi.org/10.1097/opx.0b013e31804f81a9
  6. Loewenstein, J. I., Palmberg, P. F., Connett, J. E., & Wentworth, D. (1985). Effectiveness of a Pinhole Method for Visual Acuity Screening. Archives of Ophthalmology, 103(2), 222–223. https://doi.org/10.1001/archopht.1985.01050020074024
  7. Marmamula, S., Keeffe, J. E., Narsaiah, S., Khanna, R. C., & Rao, G. N. (2014). Population?based assessment of sensitivity and specificity of a pinhole for detection of significant refractive errors in the community. Clinical and Experimental Optometry, 97(6), 523–527. https://doi.org/10.1111/cxo.12172
  8. Rosenfield, M. (1995). Repeatability of Subjective and Objective Refraction. https://www.semanticscholar.org/paper/Repeatability-of-Subjective-and-Objective-Rosenfield-Chiu/640217123ae5db6a64dd4f56295b808d16cc0cca

Figure 1

Figure 2

antarmuka fokus mahjong daya pengguna aktifaws grid serasi mahjong dasar tahapan terjagaaws jejak mekanisme mahjong arah fase lanjutanaws kajian wild berantai mahjong interaktif analitisaws kesesuaian persentase layanan mahjong seluler lanceraws pendalaman persentase mahjong gerak wild mutakhircorak langka mahjong tumbuh perlahan berubahgerak mahjong adaptasi mekanisme pemakai sekarangnalar scatter mahjong malam putaran ekstratempo mahjong kaitan mekanisme keadaan terkinialur permainan mahjong cepat scatter wilddalam hitungan detik scatter wild mahjongmenyatukan naluri pola scatter hitam mahjongmomen mahjong permainan berbalik arahmomen singkat mahjong dinamika permainanperpaduan insting pola scatter hitam momentperubahan drastis mahjong ways scatter wildscatter wild mahjong datang polasekejap berubah scatter wild mahjong wayssensasi baru mahjong lebih scatter wildenergi scatter emas irama reel mahjongevolusi reel mahjong balutan mistisintervensi cepat emas momentum lamakemunculan mendadak naga emas mahjongketika scatter naga emas aktif mahjongnaga emas muncul arah spin mahjongnaga emas ritme mahjong ways berubahrahasia rtp tinggi balik scatter hitamsaat scatter naga emas alih irama reelscatter hitam kunci lonjakan rtp mahjonge5 scatter wild memberikan sentuhan baru di setiap spin mahjong ways 2e5 scatter wild menghidupkan suasana permainan mahjong ways 2e5 scatter wild mengubah pola permainan mahjong ways 2 secara signifikane5 setiap putaran mahjong ways 2 terasa berbeda dengan scatter wilde5 strategi adaptif berbasis analisis rtp hariane5 strategi berbasis data dan algoritma untuk analisis momentume5 strategi berkembang berkat data rtp hariane5 strategi memahami algoritma untuk mengidentifikasi momentum ideale5 strategi membaca pola algoritma demi menangkap momentum optimale5 strategi modern mengandalkan evaluasi rtp hariane5 strategi responsif dengan dukungan evaluasi rtp hariane5 strategi terukur dengan analisis rtp hariane5 struktur scatter dan wild terlihat jelas berkat analisis sistem moderne5 tanpa disadari kombinasi ini sering mengarah ke scatter di mahjong wins 3e5 teknik evaluasi algoritma untuk mendapatkan momentum yang tepate5 teknik observasi sistem untuk analisis momentum yang lebih presisie5 terungkap formasi ini sering jadi awal munculnya scatter di mahjong wins 3e5 transformasi digital rtp live berkat artificial intelligence inovatife5 transformasi ritme mahjong ways 2 dipicu oleh kekuatan scatter wilde5 wajib tahu pola tersembunyi ini sering menghasilkan scattere5 applee5 bananae5 candye5 doge5 eaglee5 falcone5 geminie5 horsee5 indiae5 japananalisa pola mahjong ways rutinanalisis kinerja heuristik variansi gameanalisis pola mahjong ways hariananalisis pola mahjong ways kebiasaanera baru mahjong wins bonus optimalgebrakan bonus mahjong wins mekanisme efisieninsight pola mahjong ways rutinkajian pola mahjong ways rutinkomparasi heuristik variansi game digitalledakan bonus mahjong wins sistem efektifmahjong wins bonus sistem generasi baruobservasi pola mahjong ways harianpendekatan algoritma heuristik variansi gameperbandingan model heuristik variansi gamerahasia bonus mahjong wins sistem cerdasrangkuman pola mahjong ways harianringkasan pola mahjong ways harianstudi pola mahjong ways hariantinjauan heuristik variansi game digitaltinjauan pola mahjong ways harianalur sombol mahjong kemunculan scatterdari rtp mahjong bermain lebih efektifjejak scatter mahjong putaran tenangkejutan scatter wild simbol mahjong arahkemunculan simbol ganda membuat mahjongketika grid mahjong scatter semakin dekatketika rtp mahjong pola mulai lebih jelasketika scatter wild ritme simbol mahjongketika scatter wild titik sesi mahjong waysketika susunan simbol mahjong ritme scattermemahami rtp mahjong cara bermain lebihpergerakan simbol mahjong scatter wildpergeseran mahjong ketika scatter hadirsaat rtp mahjong terbaca baik strategisaat scatter hadir simbol mahjong bergeserscatter wild dinamika simbol mahjongstabilitas putaran mahjong pola scattersusunan baru reel mahjong scatter emassusunan mahjong wins mengandung scattersusunan simbol mahjong diam pola scatterrm menguak keunikan mahjong wins sudut pandang teknisrm cara memahami pergerakan mahjong ways tenaga ekstrarm mahjong wins standar baru industri hiburan digitalrm rahasia ketahanan mahjong ways eksis gempuran gamerm pentingnya memahami transisi level mahjong wins mendalamrm strategi mengatur tempo mahjong ways kendali permainanrm peran kecerdasan buatan mekanisme mahjong wins adilrm alasan keberhasilan mahjong ways mencuri perhatian analisrm mempelajari struktur dasar mahjong wins efisiensi putaranrm inovasi desain mahjong ways kesan bermain responsifrm teknik observasi mahjong wins jarang dibahas dampakrm cara mempertahankan fokus dinamika mahjong ways cepatrm eksplorasi fitur tersembunyi mahjong wins ritme terbaikrm mahjong ways integrasi teknologi modern keamanan nyamanrm analisis faktor pendukung mahjong wins digemari generasirm langkah efektif menyesuaikan perubahan sistem mahjong waysrm mengintip proses pengembangan mahjong wins kualitas penggunarm analisis data membantu membaca arah mahjong waysrm menemukan titik temu insting logika mahjong winsrm transformasi besar mahjong ways menghadirkan tantangan menarikmengungkap simbol langka nasib drastismisteri besar kombinasi simbol langkamisteri simbol langka keberuntungan besarsimbol langka misterius ubah hiduprahasia simbol langka nasib cepattransformasi bonus mahjong wins sistem efektifmahjong wins suguhkan bonus sistem modernsuguhan bonus efisien mahjong winsefektivitas sistem bonus mahjong winsmahjong wins hadirkan bonus sistem optimaloke76cincinbetaqua365slot gacorstc76samurai76TOBA1131samurai76 login